中文 | EN | TEL:+86-027-8757-0662 Ext. 812 ; Email: cs@healthcare-bio.com

Position:Home > Product Centre > COVID-19/SARS-COV-2 Detection Kits Product Centre

  • +86-027-8757-0662 Ext. 812 ; Email: cs@healthcare-bio.com
  • cs@healthcare-bio.com
  • Bldg. #8 Optics Valley Precision Medicine Industry Base, #9 Gaokeyuan 3rd. Rd., East-Lake High-Tech Zone, Wuhan, Hubei, China.

COVID-19 Standard Test Kit

Product Cat. No.: RP-010. / CE-IVD Registration No.: DE/CA20/IVD-Luxuslebenswelt-62/20.

 

Corona Virus Disease 2019 (COVID-19) Nucleic Acid Detection Kit 

 

[Product Name] Corona Virus Disease 2019 (CoViD-19) Nucleic Acid Detection Kit (Real-Time PCR Method).


[Packing Specifications] 100 Tests/Box.


[Intended Usage]

This kit is used for in vitro qualitative detection of nasopharyngeal swabs, oropharyngeal swabs, and alveolar lavage fluids of suspected cases of pneumonia, suspected aggregation cases, and other patients who need to be diagnosed or differentially diagnosed for new coronavirus infection. The new kit is used to detect ORF1ab and N genes.


[Detection Principle]

This kit is designed on the new coronavirus ORF1ab and the conserved region encoding the nucleocapsid protein N gene sequence published on the Global Initiative on Sharing All Influenza Data (GISAID) platform. Two pairs of specific primers and Taqman probes were used to analyze quantitatively the viral nucleic acids in the sample by using one-step fluorescence PCR detection technology.

The PCR reaction system contains primers and probes with internal standards. Through the detection of internal standards, the sample collection, extraction and reaction process are monitored to avoid false negative results.


[Product Main Components]

Kit composition

Component name

Specifications

Quantity

Main components

Reaction solution

850μL

2 Tubes

Specific primers, probes, 5 × RT PCR Buffer

Mixed enzyme solution

150μL

2 Tubes

Reverse transcriptase, DNA polymerase, dNTP, Mg2+

Negative control

400μL

1 Tube

Pseudovirus containing internal control fragment

Positive control

400μL

1 Tube

Pseudovirus containing target fragments and internal control fragment

Instructions Manual

-

1 copy

-


Required but not provided: Nucleic acid extraction or purification reagents.

Note: 1. The components of different batch kits cannot be used interchangeably.

          2. The positive control and the negative control need to be used in the nucleic acid extraction.

 

[Storage Conditions & Validity]

1. The kit should be stored frozen at -20℃±5℃ and protected from light; the expiration date is 6 months; the production date and expiration date are shown in the outer packaging box.

2. Avoid repeated freezing and thawing of the kit and the number of freezing and thawing shall not exceed 7 times.

3. After opening, it should be stored at -20℃±5℃ and protected from light. The number of bottle openings shall not exceed 7 times, which will not affect the use within the validity period.


[Applicable Instruments]

1. This kit is suitable for ABI7500 quantitative fluorescence PCR instrument.

2. For other models not listed, relevant experiments have not been conducted with this kit. If the user needs to apply any other instrument to carry out the detection of this reagent, please contact our Technical Department for relevant support.

NB: The other devices can include QuantStudio (ThermoFisher), Applied Biosystems, Bio-Rad (CFX96 Touch 1000, iQ5 iCycler), Agilent (MX-3000, Aria-Dx), MGITech, LighCycler 480II & Z480 (Roche), M2000RT Abbott, Pentabase Gentier 48E, SLAN-96S, HIMEDIA INSTA Q96 LA1012, HIMEDIA INSTA Q48 LA1023, Mygo Pro, Bioer 9600, Optimum Gene (MCL-Q2KB, MCL-Q2KC), Rotor-Gene (6000, 3000, Q6Plex), LineGene 9600, DT-Lite, DT-96, DT-Prime, Biobase FQD-96A thermocycler series and other quantitative fluorescence PCR platforms with FAM, VIC/HEX and Cy5 fluorescence detection channel.

 

[Sample Requirements]

1. Samples are from nasopharyngeal swabs, oropharyngeal swabs, etc. The operation method is as follows:

     i) Nasopharyngeal swabs: The operator gently rotates the wet sterile swabs parallel to the angle of the upper jaw from left to right and from one nostril to the inner nasopharynx of the nasal canal. Generally, when the swab is inserted with resistance, the swab will stay for 2-3s and then slowly turn it to exit.

     ii) Oropharyngeal swab: The operator holds the tongue with the tongue depressor in one hand and the root of the sterile swab in the other. Scrape the secretions from both sides of the tonsils and the pharyngeal wall posterior separately.

     iii) Place the nasopharyngeal swab or the oropharyngeal swab into the centrifuge tube containing 1.0mL physiological saline solution for later use.

2. Alveolar lavage fluid sample: Use a sterile syringe to take the sample and place it in a centrifuge tube for examination.

3. Cross-contamination between samples should be avoided.

4. The samples should be tested in time after collection, or stored at -20±5℃ for testing, and kept below -70℃ for long-term storage.


 [Testing Method]

1. Reagent preparation: (Reagent preparation area)

In each PCR reaction, a positive control and a negative control are tested simultaneously, and each sample needs to be tested for the ORF1ab region, N gene, and internal standard.

1) Remove the kit from the refrigerator, equilibrate at room temperature, fully dethaw each component, vortex to mix and then centrifuge instantly.

2) Calculate the number of reactions required for the current experiment n (n = number of samples + negative control + positive control), and mix the reagents according to the preparation method of the reaction system in Table 1 and then divide into 20μL/well PCR reaction tube. Transfer the PCR reaction tube to the sample preparation area, and put the remaining reagents back to -20℃±5℃ and keep away from light.


                                         Table 1: Reaction system preparation method

Composition

Volume

Reaction solution

17μL

Mixed enzyme solution

3μL

Total volume

20μL

 

2. Sample preparation: (Sample preparation area)

1) RNA extraction:

·         It is recommended to use commercially available nucleic acid extraction kits to extract RNA sample for PCR detection. Extract  

           according to the operation steps of the Extraction Kit instructions.

·         The positive control and negative control of this kit are part in the extraction. 

2) Loading/Sample addition:

a. Take out the reagent prepared in the reagent preparation area and centrifuge at low speed for 10 s.

b. The RNA of the sample to be tested, the positive control and the negative control after treatment are respectively added into the PCR reaction tubes at the volume of 5μL/well.

c. Cover the PCR reaction tubes, record the sequence of template loading, and centrifuge at low speed for 10 s.

d. Transfer the PCR reaction tubes to the nucleic acid amplification area for operation.

Note: Avoid contamination during RNA sample extraction and loading. If the extracted RNA template cannot be immediately tested, it is recommended to store at or below -70oC.

 

3. PCR: (Nucleic acid amplification area)

1) Boot, preheat and check the performance of the instrument.

2) Take the PCR reaction tubes prepared in the sample preparation area and place them in the corresponding positions in the sample tank of the machine (be careful to check whether the reaction tubes are tightly closed before starting the machine to avoid aerosol pollution of the instrument and the environment due to PCR products leakage) and record the placement order.

3) Set the nucleic acid amplification relevant parameters in the machine according to Table 2 and carry out PCR amplification.

 

                           Table 2: Instrument nucleic acid amplification related parameters

System

Set the reaction system to 25μL

 

Signal acquisition

      Select FAM and VIC channels to detect viral ORF1ab and N genes.

      Select Cy5 channel to detect internal control.


 

 

PCR reaction conditions

Stage

Conditions

Number of cycles

   Reverse Transcription

50℃: 15min

1

      Pre-denaturation

95℃: 15min

1

 

PCR

94℃: 15s

 

45

55℃: 45s

(Collect fluorescence signal at the end of this step)

Note: ABI series fluorescence quantitative PCR instrument does not choose ROX calibration, and select <None> for quenching group. 


4. Reading of test results

For data analysis, select ΔRn VS Cycle and Linear mode; select Auto Threshold and Auto Baseline, view the fluorescence curve, and obtain the fluorescence quantitative curve and Ct value in the fluorescence quantitative PCR instrument.


5. Quality control procedures

1. Positive control: FAM and VIC signals have obvious amplification curves at the same time, Ct value ≤32, Cy5 signal has or does not have amplification curve.
2. Negative control FAM and VIC signals have no rising curve; Cy5 signal has a significant amplification curve, Ct value ≤32.
3. The above requirements shall be met at the same time in the same experiment, otherwise, the experiment is invalid and needs to be carried out again.


6. Interpretation of results

1. According to the below table, determine the detection results of the new coronavirus.


Test results


Test results and interpretation

N gene
(FAM channel)

ORF1ab
(HEX/VIC channel)

Internal Control

(Cy5 channel)

+

+

+

Positive

+

+

-

Positive

+

-

+

Suspected positive, repeat test

-

+

+

Suspected positive, repeat test

-

-

+

Negative

-

-

-

Invalid

       (+): Presence of amplification S shape curve; (-): Absence of amplification S shape curve


2. If the test sample detects a typical S-type amplification curve in the FAM and HEX/VIC channels and the Ct value is ≤40, and there is a typical S-type amplification curve in the Internal Control channel (Cy5), the sample is judged as COVID-19 positive. 

3. If the test sample has no amplification curve or Ct value >40 in the FAM and HEX/VIC channels, and there is a typical S-type amplification curve in the Internal Control channel (Cy5), the sample is judged as COVID-19 negative.

4. If the test sample only has a Ct value ≤40 in a single channel of the FAM or HEX/VIC channel, and there is a typical S-type amplification curve in the Internal Control channel (Cy5), the results need to retest. If the results repeated are consistent, then the sample is judged as COVID-19 positive.

5. If no typical S-type amplification curve (No Ct value) or Ct value >40 is detected in the FAM, HEX/VIC, and Cy5 channels of the test sample, it means that there is a problem with the quality of the sample or a problem with the operation, so the result is invalid. The cause should be found and eliminated. After, sample should be again collected and the test repeated (if the test result is still invalid, please contact our technical service at cs@healthcare-bio.com).


[Positive Value Determination]

1. Positive result: The fluorescent quantitative curve has amplification curve, Ct value ≤40.
2. Negative result: The fluorescent quantitative curve has no amplification curve or Ct value> 40.


[Test Results Interpretation]

1. Negative and positive controls shall be tested in each experiment and the test result can be determined only when the controls meet the quality control requirements.
2. When FAM and VIC test channels are positive, Cy5 channel result may be negative.
3. When the internal standard result is negative, if the FAM and VIC test signals of the test sample are also negative, the test result of the sample is invalid, and the cause should be found and eliminated, and the experiment repeated for this sample.


[Product Performance Index]

1. Specificity: α-interferon, Zanamivir, Ribavirin, Oseltamivir, Paramivir, Lopinavir, Ritonavir (Litonavir) Abidor (Abidol), Levofloxacin, Azithromycin, Ceftriaxone , Meropenem, Tobramycin, Histamine hydrochloride, Phenylephrine (Benzolin), Oxymetazoline (Hydroxymezoline), Sodium chloride were selected to test their interference with this kit and the results did not reveal any interfere with this kit.

2. Cross-reaction: Other pathogens that are similar to the new coronavirus 2019 (COVID-19) species or cause similar symptoms (new H1N1 influenza virus (2009), seasonal H1N1 influenza virus), influenza A H3N2, H5N1, H7N9, influenza B Yamagata , Type B influenza Victoria, respiratory syncytial virus type A, respiratory syncytial virus type B, parainfluenza type I, parainfluenza type II, parainfluenza type III, Adenovirus 1, 2, 3, 4, 5, 7, 55 Types, Enteroviruses A, B, C, D, Epstein-Barr virus, Measles virus, Human cytomegalovirus, Rotavirus, Norovirus, Mumps virus, Varicella-zoster virus, Mycoplasma pneumoniae, Chlamydia pneumoniae, Legion Bacteria, Pertussis, Haemophilus influenzae, Staphylococcus aureus, Streptococcus pneumoniae, Streptococcus pyogenes, Klebsiella pneumoniae, Mycobacterium tuberculosis and human genomic DNA did not cross-react.

3. Precision: The coefficient of variation Intra-batch / Inter-batch, Intra-day / Inter-day among different operators shall not be higher than 5.0%.

4. Coincidence rate of negative positive controls: The coincidence rate of 7 positive controls and 10 negative controls is 100%.

5. Minimum detection limit: The minimum detection limit of this kit is 500 copies/mL. 


[Test Method Limitations]

1. The test results of this kit are for clinical reference only. The clinical diagnosis and treatment of patients should be considered in combination with their symptoms/signs, medical history, other laboratory tests, and treatment response.

2. The improper operation of the sample during the collection, transportation, storage and nucleic acid extraction process can easily cause RNA degradation and lead to false negative results.

3. When the detected nucleic acid concentration in the sample is less than the minimum detection limit, false negative results may occur.

4. If cross-contamination occurs during sample collection and preparation, it is easy to lead to false positive results.

5. Some infected people have a large number of dead virus in their samples due to the taking of antiviral drugs. At this time, there may be strong positive results detected by this kit and negative results detected by the culture method. In such case, the recent medication situation of the tested person shall be inquired.

6. Variations of the targeted sequence of the virus to be tested or other cause of sequence changes may lead to false negative results.

7. For a new type of virus, the most suitable sample type and the best sampling time after infection may not be confirmed. Therefore, the possibility of false negative results will be reduced if the samples are collected in the same patient in different times and multiple parts.

 

[Precautions]

1. Laboratory management shall be strictly implemented in accordance with the management specifications of the “Administrative Measures for Clinical Gene Amplification Laboratory of Medical Institutions” promulgated by the General Office of the Ministry of Health.

2. Laboratory personnel must have professional training and have certain experience.

3. The experimental process should be carried out in zones (reagent preparation zone, sample processing zone, nucleic acid amplification zone). Special instruments and equipment should be used at each stage of the experiment operation, and supplies should not be used in each stage. There should be strict requirements to minimize cross-contamination.

4. Experimental consumables (such as centrifuge tubes, suction heads, etc.) should have reasonable cleaning and quality inspection procedures to prevent contamination from causing false positive results or amplification reaction inhibitors to cause false negative results.

5. The instrument and its supporting power supply system should be checked before use to ensure the normal operation of the reagent after the reagent is out on the machine.

6. The suction heads used in the experiment shall be directly put into the waste tank containing 10% sodium hypochlorite and discarded together with other waste products.

7. The workbench and various experimental items are often disinfected with 10% sodium hypochlorite, 75% alcohol and UV lamps.

8. The Real-Time PCR machine requires frequent calibration and sample plate wells (holes) cleaning.

9. To prevent fluorescence interference, avoid direct contact with the octaplex PCR reaction tubes and tubes covered by hands.

10. The positive control in this kit is not contagious and will not cause harm to the human body. However, it is recommended to treat it as a potentially infectious substance when using it.

11. The test samples involved in this kit should be considered as infectious substances, and their handling must meet the relevant requirements of the General Guidelines for Biosafety of Microbiology and Biomedical Laboratories of the Ministry of Health and the Medical Waste Management Regulations.


                                    PRODUCT KEY FEATURES

Product Name

Corona Virus Disease 2019 (CoViD-19) Nucleic Acid Detection Kit

Method

Real-Time PCR Method

Gene Target

ORF1ab and N genes

Sample Type

Upper respiratory tract specimen: Nasopharyngeal fluids, Nasal swabs; Lower respiratory tract specimen: Sputum, Bronchoalveolar lavage, Tracheal aspirate.

Verified LOD (Copies/mL)

500

Average Ct

(Lowest dilution 10/10)

35-36

Clinical Sensitivity

100% (95%CI:93, 100)

Clinical specificity

99% (95%CI:95, 100)

Manufacturer Recommended

Ct Cut-off

Positive: ≤40

Negative: >40

Thermocycler Run Time

110 minutes

PCR Platform (Thermocycler)

Most 3 channels thermocycler devices such ABI7500 & QuantStudio (ThermoFisher), 

Applied Biosystems, Bio-Rad (CFX96 Touch 1000, iQ5 iCycler), Agilent (MX-3000, 

Aria-Dx), MGITech, LighCycler 480II & Z480 (Roche), M2000RT Abbott, Pentabase Gentier 48E, SLAN-96S, HIMEDIA INSTA Q96 LA1012, HIMEDIA INSTA Q48 LA1023, Mygo Pro, 

Bioer 9600, Optimum Gene (MCL-Q2KB, MCL-Q2KC), Rotor-Gene (6000, 3000, Q6Plex), 

LineGene 9600, DT-Lite, DT-96, DT-Prime, Biobase FQD-96A thermocycler series and 2

other quantitative fluorescence PCR platforms with FAM, VIC/HEX and Cy5 channels.




Corona Virus Disease 2019 (CoViD-19) Nucleic Acid Detection Kit (Real-Time PCR Method).

Experiment Process